Optopol
Configure Dicom Exports
The recommended settings for macula scan are:
Retina 3D, horizontal b-scans (no fundus image should be exported with b-scans)
Retinal area of interest – from 6 till 12 mm.
Scan Width - 1024
Scan Height - 944
Altris AI analyses the central sector of the retina and does not provide optic nerve analysis
Step 1
- From the Main PC screen Enter the Password for SOCT log in.
- Once entered click the Set up button to enter the set up menu.
Step 2
- In Set Up Menu click the preferences tab
- Click Output Settings
Step 3
- In ‘Output Settings’, Scroll down to the ‘Tomogram Series DICOM’
- Click to Select
- Select Edit
Step 4
- If Selected untick the ‘Add Fundus Image’ box
- Select the ‘Image Compression’ Tab
- Within Drop Down Box Select/highlight JPEG (Lossy) option
- All settings are complete please click OK back to main screen (3 times)
Step 5
Image Transfer
- Select Patient
- Select Scan
- Click’ Output’
Step 6
- Scroll Down until Altris Dicom can be selected (output name)
- Click OK
- Wait while the file is saved (This file is saved to the Altris Folder on the DeskTop)
- When complete the screen will auto refresh back to previous screen.
- File is then stored in Altris Folder on Desktop for file upload.
Updated about 1 month ago